1. Signal Compatible with IV sets of any standard;
2. High accuracy and eliminates false alarm with the unique design of drop sensor;
3. Three work modes: Drop counting control, volume control and time control mode;
4. 8 hours build-in battery life in high capacity;
5. Special light for night operation;
6. Three levels of occlusion pressure;
7. Three levels of alarm volume;
8. Switch ml/h and drop/min freely;
9. Self-correcting the occlusion alarm according the soft hardness of the IV sets;
10. Waterproof and anti- sticking design of the drop detector;
11. Bolus function.
SPECIFICATIONS:
Controls: Drop counting and adjusting the delivery rate automatically
Pump Type: Peristaltic finger
Applicable IV Sets:
Compatible with IV sets of any standard(including lucifugal IV sets)
Accuracy: 25drop/min: 2%, >25drop/min: 3% (With the drop counting control mode)
5% (With the rate control mode after the correct calibration)
Delivery Rate Range:
1~400 drop/min(1drop/min step)
1.0ml/h~1200ml/h (1.0ml/h~99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step)
Delivery Time Range: 1min~1999min (1 min step)
Volume Limit Range: 0~9999ml (1 ml step)
Total Volume Range: 1~9999ml (1 ml step)
Rapid injection: 600ml/h
KVO Rate:
When the delivery rate: 10ml/h, KVO is 1 ml/h. Otherwise KVO is 3ml/h.
Alarms:
Air-in-line, Infusion complete when limited volume has been set, Empty container, Occlusion, Door is open, waitting time Over, No drop been detected, Operational errors, Low battery, other device failures.
Power Supply:
AC power supply: AC 100 to 240V, 50/60Hz;
Internal LI Battery: 11.1V, 2000mAh. 8 Hours of continous operation on a new fully charged battery with the rate of 20 drops/min.
Power Consumption: 25W
Safety level: ClassI, Type BF
Dimensions: 130mm(L) X200mm(W)X230mm(H)
Net Weight: Approximately. 2.5kg
Operating Conditions: Ambient temperature: 5 to 40Cent Relative humidity: 20 to 90%
Storage Conditions: Ambient temperature: -10 to 55cent Relative humidity: 10 to 90%
Accessories: AC power line, Operational manual