We can supply different size screen patient monitor, the screen size range from 15''/12''/10''/8''/7''
SHM-3010
10.1" active matrix TFT LCD Screen
ECG, SpO2, NIBP, RESP, TEMP, HR and PR
Optional Dual-IBP, EtCO2, Built-in thermal recorder, Built-in battery, Pacemaker pulse rejection
PVC analysis
Defibrillator overload protection
Electrosurgery protection
Pacemaker detector
HRV and 13 kinds of Arrhythmia analysis S-T segment analysis
300 seconds ECG waveform review
Maximum 72-hour graphic and tabular trends of all parameters and 72-hour data records, 100 items alarm event records
Adjustable 3-level audible and visual alarm
Application from adult to neonate
Support wired compatible with Safeheart central monitoring system
Technical Specifications
Power: 100~250V AC, 50/60 Hz
Battery:
Lead-acid recharge battery
Working more than 100 minutes when full charge
about 5 minutes working after first low-power alarm
Battery charge time is not more than 12 hours
ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, AVR, AVL, AVF, V
Gain: X 250, x 500, x 1000, x 2000
HR range: Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which is greater
Resolution: 1 bpm
Sensitivity: >200 UV
Input resistance: >5M Ohm
CMRR: Diagnosis mode: >90 dB
Monitor mode: >100 dB
Operation mode: >100 dB
Electrode offset potential: ± 300mV
Pacemaker Pulse Detecting: YES
ECG signal range: ± 8 mV
S-T detection: YES,
Arrhythmia analysis: YES
Alarm: YES, audible and visual alarm, alarm events recallable
RESP
Technique: Trans-thoracic impedance (RA-LL or LA-LL)
Measurement range: Adult: 6 ~ 120 BrPM
Pediatric/neonate: 6 ~ 150 BrPM
Resolution: 1 BrPM
Accuracy: ± 2 BrPM
Apnea alarm: 10 ~ 40 s
SpO2
Measurement range: 0 ~ 100%
Resolution: 1%
Accuracy: 70 ~ 100%: ± 2%
0 ~69%: Unspecified
Pulse Rate
Measurement range: 0 ~ 254 bpm
Resolution: 1 bpm
Accuracy: ± 2 bpm
TEMP
Transducer: YSI/CYF series
Channel: Dual-channel
Measurement range: 0 ~ 5O
Resolution: 0.1
Accuracy: ± 0.1
NIBP
Measurement mode: Manual/auto pulse wave oscillation
Measurement interval time: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480min
Continuous measuring time: 5 min
Measurement range:
Adult:
Systolic: 40 ~ 270 mmHg
Diastolic: 10 ~ 215 mmHg
Mean arterial: 20 ~ 235 mmHg
Pediatric:
Systolic: 40 ~ 200 mmHg
Diastolic: 10 ~ 150 mmHg
Mean arterial: 20 ~ 165 mmHg
Neonate:
Systolic: 40 ~ 135 mmHg
Diastolic: 10 ~ 100 mmHg
Mean arterial: 20 ~ 110 mmHg
Accuracy: The mean error shall be less than ± 5 mmHg, the standard deviation shall be less than ± 8 mmHg
Overvoltage protection: YES
Recorder(optional)
Width: 48mm
Speed: 25mm/S
Wave: 2
Type: 8s, 16s, 32s real-time record
IBP(optional)
Channel: 2 channels
Transducer sites: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART: 0 ~ 300 mmHg
PA: -6 ~ 120 mmHg
CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
P1/P2: -10 ~ 300 mmHg
Press Sensor Sensitivity: 5 UV/V/mmHg
Resolution: 1 mmHg
Accuracy: ± 2% or ± 1mmHg, which is greater
Actualization interval: about 1s
EtCO2(optional)
Measurement Mode: Side-stream or Mainstream
Measurement range: 0% ~ 13%
Resolution: 1 mmHg
Accuracy: 2 mmHg @ <5.0% CO2 (at ATPS)
Apnea alarm: YES