Power output: 3.5KW
X-ray Tube: XD3-3.5/100
Commute Method: Full wave commutation
Tube Voltage: 40-90KV
Tube current: 40-90KV
40-90KV 30mA
40-80KV 50mA
Radiograph Time: 40-90KV 15mA 0.1-6.3s
40-90KV 30mA 0.1-6.3s
40-80KV 50mA 0.1-2s
Power supply: 220V ± 10% 50Hz
Inner-resistance≤ 1.0 Ω 4KVA
Main technical parameters:
1. Power conditions:
Power voltage: Single-phase A. C 220V± 22V
Power frequency: Sine wave 50Hz± 1Hz
Power capacity: 3KVA
Internal resistor of the power supply: ≤ 1Ohm
Fuse of the power supply: 15A/220V
2. Maximum rated capacity of Radiography:
Tube current: 50mA
Tube voltage: 80kv
Duration: 2s
3. Radiography conditions:
Tube voltage: 40-90KV 9 shifts for adjustment, 5KV for each shift.
Tube current: 30mA and 50mA shifts.
4. Maximum output power: Intermittent mode: 3KVA
5. Nominal electric power: P=0.74× 80KV× 50mA=2.96KW
Features:
1. Radiation protection of the X-ray machine shall be in line with GB9706.12.
2. Total filtration of the X-ray machine shall be 2.5mmAL, wherein 1mm is the additional filtration piece.
3. The distance from the focal spot to the skin is 20cm.
4. A varied vision field single-page beam limited device is installed at the window of the X-ray generator, and the maximum radiography specifications 330mm× 330mm can be adjusted 1M away.
5. A vision field indicator lamp with power supply of 12V is provided for the beam limited device.
6. The frame can move forwards, backwards, leftwards and rightwards.
7. The brake is composed of the brake and brake release. If the integral fixing is required, step on the brake pedal leftwards or rightwards; When releasing the brake, step on the brake to release the pedal.
8. The long arm employs spring balance. When the X-ray generator is hung, the X-ray generator can be stay at any specified position.
9. Rotating device of the X-ray generator
The X-ray generator can rotate with the frame. When the X-ray generator is at the horizontal and vertical position, it rotates ± 90° Around itself.