Specifications:
--- 23 parameters. Throughput: >60T/H
--- 10.4'' TFT LCD
--- High accuracy technology for blood processing, only 9.6µ L blood sample
--- Storing 40 thousand samples. Operated by mouse & keyboard.
Technical Specifications:
Function: Automated WBC 3-differential, 23 parameters with 3 histograms(WBC, RBC, PLT)
Principal: Electronic impedance, SFT method for hemoglobin
Testing Channel: Two channels testing, WBC/HHGB channel and RBC/PLT channel
Parameters: WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, PCT, RDW-CV, RDW-SD, MPV, PDW, P-LCR and WBC, RBC, PLT three histograms.
Throughput: >60T/H, 24 hours switch on, automatic sleep and wake up function
Sampling Mode: Optional venous blood, anticoagulant blood and prediluted blood
Sample Volume: Whole blood ≤ 9.6µ L, capillary blood ≤ 9.6µ L, prediluted capillary blood ≤ 20µ L, the sample volume for repeating two times testing and avoiding second sampling
Working Mode: Automatic sampling, detecting automatic testing, furnished with dilutor, no need diluting outside. Automatic dispensing lyse and making second diluting automatically.
Block-clearing: Soak and washing back and forth. Real time monitoring function of high pressure burning and automatic block-clearing. Boast the function of monitoring incomplete blockage
Alarm: Automated system diagnosis, self testing, fault alarm and reagent shortage alarm
Automated Maintenance: Automated washing during power on and power off, reducing the blockage rate.
Storage: Adopting 2G SD card memory. Automatic storing 40 thousand samples with all parameters and 3 histograms. Combined search mode as per many conditions
Control Mode: Boasts management system for quality control, nine kinds of control files. Mean value, SD value, CV value are provided. Automatic drawing L-J graph. Provide matched blood control
Calibration Mode: Automatic calibration and manual calibration
Reference Value Setting: Man, Woman, Children, Baby and General etc.
Performance: Repeatability Error Linearity Range.
WBC <2.0% 0.0~99.9× 109/L
RBC <1.5% 0.00~9.99× 1012/L
HGB <1.5% 0~300g/L
MCV <0.4% 40~150fL
HCT <1.0% 0.0~100.0%
PLT <4.0% 0~999× 109/L
Carry Over: RBC, WBC, HGB<0.5%, PLT<1.0%.
Operating Interface: Colorful Windows interface, no need connecting computer. Operating by graphic button with mouse and keyboard
Display: Big size 10.4'' TFT LCD, all parameters and histograms displayed in the same screen
Print Mode: Internal high-speed thermal printer. Optional connection with specified external printer. English printouts
Reagents System: Open reagent system. No cyanogen reagents. Provide original reagents.
Connection Port: USB, RS-232, network interface
Power: Maximum 150VA
Environment: Temperature 10C-35C. Relative humidity: 10%-90%
Power Supply: 220(1± 10%)V, (50± 1)Hz. 90V-260V, 50-60Hz broad range power available for different Countries
Dimension: 520mm× 460mm× 420mm
Weight: 23Kg