MHJ-IIIBl is a universal mulit-functional anesthesia Machine incorporating anesthesia, respiratory and Monitoring, In design, it improves the reliability, safety And combination. Also consideration has been taken to the Connection between anesthasia machine and all kinds of Monitors, extending its functions. This machine is equipped With Moedl SC-M3A1 pneumatic-driven and pneumatic-Controlled anesthesia ventilator, C02-1V breathing circle System and DZFG high-precision vaporizer, suitable for Closed, semi-closed and semi-opened inhalation anesthesia And non-inhalation anesthesia in all levels hospitals.
O Equipped with SC-M3A1 pneumatic-driven and pneumatic Controlled anesthesia ventilate without power.
O Flowmeter with ORC system to ensure oxygen concentra Tion not less than25%.
O N20 cut-off device.
0 02 deficiency alarm.
O SC-M3A1 anesthesia ventilator, with easy operation and Low-air consumption.
O Top plate Intended for all kinds of monitors.
O Gas Supply: 0.27-0.55Mpa.
O Adjustment range of flow: 02: 0.1-15L/min, N20: 0.05-10Umin.
O RC: System: 02 concentration; 25 % (02+N20).
0 02 deficiency alarm: Audible alarm occurs when 02 pres Sure<0.2Mpa.
0 02 flush: 35-75Uniifl(02 pressure 0.45Mpa)
N20 is out off when 02 pressure≤ O. 1Mpa
O Vaporizer: 0.2-5VOL%
O Adjustment range of APL: 0.5-7kPa
O SC-M3A1 anesthesia ventilator:
Freq: 6-40bpm
MV: 3-18L/min
1: E: 1: 1.5(fixed)
O Safety: Working pressure: ≤ 6KPa
O Monitor: Airway Pressure Tidal Volume, MV, Gas Resoutce pressure.